The Elisa Microplate Reader AMR-100/AMR-100T is a high-quality light absorption microplate reader based on a filter, with a wavelength range of 340 nm~750 nm, suitable for scientific research and clinical applications. 7-inch touch screen color LCD display, no keyboard required, easy to use. Data and programs can be saved in the instrument or exported via a USB stick. The plate reading speed is fast, which can realize fast measurement, and provide measurement results with high accuracy and good reproducibility.
Features
- High-resolution 7-inch color touch screen, easy to operate, no keyboard required, easy to use;
- Multiple software configuration, which can be used as a single machine or connected with a computer, and the results are exported in real time;
- Absorbance range: 0.0~4.000 Abs, meeting different measurement requirements;
- 8-position filter wheel, standard 4 filters, optional other filters;
- The built-in software can provide instrument control and data analysis, and can be directly connected to a U disk;
- The detection speed is fast, and the whole 96-well microplate detection can be completed within 6 seconds.
- High-resolution 7-inch touch screen, simple and intuitive operation, no keyboard required; Visual layout, convenient and practical;
- The microplate reader is equipped with the standard control and data analysis software Readerlt-I, which is convenient and quick for data detection;
- Powerful data analysis function and excellent result report, not only can run the analysis independently, but also run the analysis on the computer.
Technical Specifications:
Model | AMR-100 | AMR-100T |
Display | 7 inch touch screen | 7 inch touch screen |
Light source | Quartz-halogen lamp 6 V 10 W | Quartz-halogen lamp 6 V 10 W |
Wavelength range | 340 nm~750 nm | 340 nm~750 nm |
Filter | 8-position filter wheel, standard 4 filters: 405nm, 450nm, 492nm, 630nm, other filters are optional | 8-position filter wheel, standard 4 filters: 405nm, 450nm, 492nm, 630nm, other filters are optional |
Read-out range | 0~4.000Abs | 0~4.000Abs |
Linearity | [0.000~3.000 Abs] R²≥0.995 | [0.000~3.000 Abs] R²≥0.995 |
Resolution | 0.001Abs | 0.001Abs |
Wavelength accuracy | ≤±2 nm | ≤±2 nm |
Absorbance repeatability | [0,3 Abs) CV≤0.3 % [3,4 Abs) CV≤1 % | [0,3 Abs) CV≤0.3 % [3,4 Abs) CV≤1 % |
Absorbance stability | ≤0.005Abs [0,3Abs) ≤2.0% [3,4Abs) | ≤±0.005A [0,2Abs) ≤0.3% [2,3Abs) ≤2.0% [3,4Abs) |
Absorbance accuracy | ≤±0.005Abs [0,2Abs) ≤±0.01Abs [2,3Abs) ≤±1.5% [3,4Abs) | ≤±0.005A [0,2Abs) ≤±1% [2,3Abs) ≤±1.5% [3,4Abs) |
Sensitivity/detector | ≥0.01 A/photodiode | ≥0.01 A/photodiode |
Test speed | 6 s/96-well, fast measurement mode; single wavelength <15 s/96 well, dual wavelengths <28 s/96 well (normal mode) | 6 s/96-well, fast measurement mode; single wavelength <15 s/96 well, dual wavelengths <28 s/96 well (normal mode) |
User interface | Built-in software, touch screen input, external mouse | Built-in software, touch screen input, external mouse |
Storage | 1000 test records | 1000 test records |
Port | 3 USB ports, for PC, printer and USB-disk | 3 USB ports, for PC, printer and USB-disk |
Power supply | AC100-240 V,50-60 Hz,2 A | AC100-240 V,50-60 Hz,2 A |
Dimension (W×D×H) | 295×440×225 mm | 295×440×225 mm |
Weight (kg) | 10 | 11 |
Temperature accuracy | – | 37 ℃:±0.5 ℃ |
Temperature uniformity | – | 37 ℃:±0.5 ℃ |
Incubator | – | RT+4 ℃~50 ℃ |