Features
- Direct detection of high concentration samples without dilution, maximum detection concentration up to 4500 ng/μL (dsDNA)
- Android system, 7-inch capacitive touch screen, optimized APP software
- Newly designed OD600 optical path detection system, new cuvette mode, for bacteria concentration detection
- High resolution CCD array detector, 5 s can complete the detection and display the results
- Long life pulse xenon lamp light source
- The test data is transferred to the computer via USB for easy sorting and analysis
- The built-in printer can print the report directly
Technical Specifications
Model | Nano-300 |
Wavelength range | 200-800 nm |
Minimum sample size | 0.5-2.0 μl |
Path length | 0.2mm 1.0mm |
Light source | Xenon flash lamp |
Detector Type | 2048-Linear CCD array |
Wavelength accuracy | 1 nm |
Spectral resolution | ≤3 nm (FWHM at Hg546 nm) |
Absorbance precision | 0.003Abs |
Absorbance accuracy | 1 % (7.332Abs at 260 nm) |
Absorbance range | 0.04 – 90A |
Detects nucleic acid up to | 2~4500 ng/μl (dsDNA) |
Measurement time | <5 S |
Data output | USB |
Sample pedestal materia | Aluminum alloy and Quartz fiber |
Operating voltage | DC 24V 2A |
Operating power | 25 W |
Standby power | 5 W |
Dimension (WxDxH)mm | 210×268×181 |
Weight | 2.8kg |
Software compatibility | Android system |
OD600nm measurement | |
Light source | LED |
Wavelength range | 600±8 nm |
Absorbance range | 0-4 A |