UV-VIS Spectrophotometer – OPTIZEN Alpha

The OPTIZEN Alpha Double Beam UV-VIS Spectrophotometer by K-Lab is vital for precise measurements in laboratories. Apex Scientific provides this advanced laboratory equipment in South Africa, designed to measure transmittance and absorbance in UV and visible light.

Utilizing a double-beam method, the OPTIZEN Alpha minimizes measurement errors, ensuring reliable results across various applications, including biotechnology and chemistry. Its user-friendly interface and rapid processing enhance efficiency. Explore our range or read our article, Precision Optical Analysis for Food, Wine, and Oil Testing – UV-Vis Spectrophotometers for more information.

The UV-VIS Spectrophotometer – Double Beam – OPTIZEN Alpha by K-Lab delivers high-precision measurements for accurate laboratory analysis. Available through Apex Scientific in South Africa, this advanced laboratory equipment measures transmittance and absorbance across ultraviolet and visible wavelengths. Consequently, researchers can perform quantitative assessments, such as concentration and purity, with consistent reliability in demanding research environments.

 

Accurate Measurements

The OPTIZEN Alpha uses a double-beam method that reduces errors common in single-beam spectrophotometers. Moreover, the reference beam compensates for light intensity fluctuations. As a result, it provides reliable and reproducible results. This feature supports precise quantitative analysis in biotechnology, environmental science, chemistry, and other applications where accuracy is critical.

 

Versatile Applications

This spectrophotometer suits a wide range of tasks. For instance, it measures DNA concentrations, evaluates chemical purity, and monitors other key parameters. Its broad wavelength range and high-resolution capability allow scientists to obtain trustworthy data consistently. Therefore, it is ideal for laboratories requiring both versatility and accuracy.

 

User-Friendly Features

K-Lab designed the OPTIZEN Alpha for convenience. The integrated interface allows fast operation and quick data acquisition. Additionally, its advanced software performs analyses up to 1.5 times faster than previous models. The compact design also maximizes workspace efficiency. Furthermore, intuitive controls reduce user error and simplify experiment management.

Overall, the UV-VIS Spectrophotometer – Double Beam – OPTIZEN Alpha remains a trusted choice for laboratories in South Africa. It ensures reliable performance across diverse scientific applications. Explore our article, Precision Optical Analysis for Food, Wine, and Oil Testing – UV-Vis Spectrophotometers, for more insights.

Photometrics SystemDouble-beam type
Light Source(s)Tungsten halogen lamp, Deuterium lamp
(Built-in light source auto interchanging motor)
DetectorSilicon Photodiode
Spectral BandwidthFixed – 1.0 nm (190 to 1100 nm)
Variable – 0.5 nm, 1 nm, 1.5 nm, 2 nm, 4 nm (optional)
Wavelength Range190 to 1100 nm
Wavelength Display (setting)0.05 nm
Spectral Bandwidth1.0 nm (190 to 1100 nm)
Wavelength Accuracy<± 0.3 nm (for entire range) <± 0.1 nm (at 656.1 nm)
Wavelength Repeatability<± 0.1 nm
Wavelength Slew Rate25,000 nm/min
Scanning Speed(Max)6,000 nm/min
Lamp interchange Wavelength340 to 410 nm (Default 370 nm)
Photometric DisplayAbsorbance: -4 A to 4 A
Transmittance: 0% to 400%
Photometric Accuracy<± 0.002 Abs at 0.5 Abs <± 0.004 Abs at 1.0 Abs <± 0.006Abs at 2.0 Abs (NIST930D/NIST1930 or equivalent)
Environmental requirementsTemperature: 15 °C to 35 °C
Humidity: 30 % to 80 %
Baseline Stability<0.0003 Abs/h (700 nm, after one hour warm-up)
Baseline Flatness<0.0005 Abs rms (190 to 1,100 nm, after one hour warm-up)
Photometric Repeatability± 0.0002Abs at 0.5 Abs
± 0.0006Abs at 1.0 Abs
± 0.0010Abs at 2.0 Abs
Stray Light< 1.00 % (at 198 nm with 1.2% KCl) < 0.05 % (at 220 nm with 0.1% NaI) < 0.05 % (at 340 nm with 0.5% NaNO2)
Photometric noise< 0.00005 Abs rms (at 700 nm)
MonochromatorCzerny-Turner Monochromator
Standard Cell HolderAutomatic Rotary type 8-position Multi-Cell Holder
Operating System (OS)Windows 10 (Embedded PC)
Display8 inch color LCD with touch screen
Control OptionsStand-alone
PC
Dimensions(W*D*H)520 × 500 × 200 mm
Power RequirementAC 100-240 V, 50/60 Hz, 140 VA
Weight14 kg
PC Software(optional) View for Windows
ConnectivityRJ45 ethernet (PC, printer)
USB A type * 2 (USB Drive, printer)
RS-232 * 1, USB B type * 1 (Debugging & FW update)
SKUDescription
UNT0002OPTIZEN Alpha Unit Spectrophotometer

Product Enquiry Handbook: UV-VIS Spectrophotometers

Peak Valley Detection Principle

Optizen View Software

Operating Manual

Product Enquiry

"*" indicates required fields

This field is hidden when viewing the form
First Name*
Last Name*