Micropipettes

What are micropipettes?
A micropipette is a common yet an essential laboratory instrument used to accurately and precisely transfer volumes of liquid in the microliter range. Micropipettes are available in single channel and multi channel variants. While the single channel micropipettes are used in labs that perform research related to molecular biology, microbiology, immunology, cell culture, analytical chemistry, biochemistry and genetics, the multichannel micropipettes are recommended for ELISA (diagnostic test), molecular screening, kinetic studies and DNA amplification.

Components of a micropipette:
Micropipettes are available in different designs and sizes. However, there are certain components that are basic and common to all the micropipettes. These include the plunger, digital display, tip cone, tip ejector and grippy. Certain micropipettes are provided with a calibration tool and a micropipette stand as an accessory.

Plunger: The plunger performs the following two functions

  1. Volume adjustment:Rotate the plunger clockwise/ anticlockwise to decrease/ increase the volume setting. A distinct click sound at every volume change ensures perfect volume setting and prevents any accidental volume change.
  2. Liquid aspiration/ dispensing: Press and depress the plunger to aspirate or dispense liquid.’

Tip Ejector:
The internal mechanism of the micropipette does not come in direct contact of the sample/liquid. Instead, a disposable pipette tip is used to draw the liquid into and dispense from the micropipette. So in order to allow the safe, effortless and quick ejection of tips, micropipettes are provided with a tip ejection system. The tips can be easily removed from the micropipette by pressing the tip ejector button.

Volume Display:
This shows the volume of the liquid to be aspirated or dispensed.

Tip Cone:
The tip cone provides fitment to the tips. A micropipette with a universal tip cone is preferred as it enhances the compatibility of the instrument with most of the standard tips.

 

Types of micropipette:
Micropipettes can be classified depending upon:

A. Working Principle: The plunger performs the following two functions:

  • Air Displacement Micropipette: This type of micropipette works on the air displacement principle. It consists of a piston which aspirates and dispenses liquid samples as the air pocket moves up and down, respectively. The internal mechanism of the pipette does not come in direct contact of the sample/liquid. Instead, a disposable pipette tip is used to draw the liquid into and dispense from the pipette.
  • Positive Displacement Micropipette: In these micropipettes, the piston comes in direct contact of the sample. The disposable tip in a positive displacement micropipette is a microsyringe composed of a capillary and a piston (movable inner part) which directly displaces the liquid.

B. Operating Mechanism:

  • Mechanical Micropipette: These micropipettes are operated manually based on a piston-shaft spring mechanism.
  • Electronic Micropipette: An electronic micropipette is mostly automated. The aspirating and dispensing of liquid is performed by the one touch buttons instead of manual plunger pressing and depressing. Electronic pipettes also often enable the user to create custom programs on the device allowing the pipettes to suit diverse application needs.

C. Number of Channels:

  • Single Channel Micropipette: A single channel micropipette is the one which has a single channel to aspirate or dispense the liquid.
  • Multi Channel Micropipette: A multi channel micropipette has multiple channels to aspirate or dispense the liquid. The commonly available multi channel micropipette variants are the 8 channel, 12 channel and 16 channel. Multichannel micropipettes reduce the workload of a single channel micropipette when working with large volumes of samples.

D. Volume/Capacity:

  • Fixed Volume Micropipette: In a fixed volume micropipette, the volume of liquid to be aspirated or dispensed remains fixed. These micropipettes are used when the same volume of liquid is to be dispensed multiple times.
  • Variable Volume Micropipette: This micropipette comes with a specific minimum and maximum volume range. The volume of the liquid to be aspirated or dispensed can be adjusted (within the instrument’s volume range) depending upon the requirement of the user.

Size and Range / Technical Specifications:
Micropipettes are available in different volumes ranging from 0.1 µl to 10,0000 µl. The commonly used variants of single channel variable volume micropipettes are listed below along with their permissible error limits as specified in the ISO 8655-2 standard.

They are sometimes referred to as P10, P20, P1000, P5000 pipettes based on the maximum volume that can be aspirated / dispensed using the pipette. For instance, a 0.5-10ul micropipette may be commonly referred to as a P10 pipette

How does a micropipette work?
Air displacement micropipettes operate by piston-driven air displacement. When the piston is pressed downwards, the air within the sleeve of the micropipette gets expelled out due to the force of which the liquid present in the micropipettes tip also gets removed.

When the piston moves upwards, a vacuum is created in the space left vacant by the piston. This causes the air from the tip to rise in order to fill the vacant space, and the tip air is then replaced by the liquid, which is drawn up into the tip.

Positive displacement micropipettes operate by piston-driven displacement. The piston in a positive displacement micropipette is in direct contact with the liquid. When the piston is pressed downwards, the liquid which is present in the sleeve of the micropipette also moves downwards and gets removed through the tip. When the piston is pulled upwards, it also draws the liquid along with it in the upward direction.

Using an Air Displacement Micropipette:
Precise measurement of liquid depends on the correct micropipette usage. The air displacement micropipettes work on the common air displacement principle. A plunger is depressed by the thumb and as it is released, liquid is drawn into a disposable tip. When the plunger is pressed again, the liquid is dispensed. In between these steps, there are several small steps which help in making the liquid dispensing process more precise.

Some micropipettes deliver the fixed volume of liquid. However, the majority are adjustable with the variable volume setting. Variable volume micropipette comes with different ranges and upper & lower limits of measurement. In such cases, error percentage may vary as per the measured liquid. Trying to dispense less than the lower value of the range will result in inaccurate liquid measurements where trying to dispense over the upper range will completely fill the tip and allow the liquid to enter into the pipette body.

Cleaning, maintenance and storage of a micropipette
The care and maintenance of a micropipette is an important routine in laboratories. Implementing a proper maintenance schedule can reduce the cost of a new purchase of this expensive equipment. Cleaning the micropipette takes time, focus and practice; otherwise the micropipette can be damaged. So, it is important to handle the micropipette carefully while cleaning. The following is a comprehensive guide for proper cleaning of micropipettes.

External Cleaning: Most of the pipette can be cleaned externally with typical laboratory cleaning agents, soaps or alcohol. To ensure the full sterilization, let the cleaning solution sit on the micropipette for 10-15 min before wiping it off.

Internal Cleaning: Cleaning the interior of the micropipette can be more time consuming because it requires full disassembly. Also, every part of micropipette will need to be cleaned properly depending on the liquid used in sampling.

  • Refer to the instructions manual for specific direction of micropipette
  • Use a cotton swab with cleaning solution and distilled water
  • Lightly grease the pistons with the lubricant provided upon purchase
  • Reassemble all parts and check to ensure the micropipette operates smoothly

Contamination Cleaning: If micropipette becomes contaminated with a known substance, there are specific cleaning steps that must be taken depending on the type of substance. Above cleaning routine will not be sufficient if the micropipette is cross-contaminated.

A well-maintained and clean pipette ensures the lab safety from hazardous solutions. It makes the pipette more accurate, reliable, long-lasting and reduces the cost of sampling.

Proper storage of micropipettes is as important as cleaning and calibrating them. The micropipette and its accessories should be stored in a clean, cool and dry place. The storage place should have a temperature ranging from -20 °C to 50 °C (from -4 °F to 120 °F) with relative humidity between 5% and 95%. Another point to be remembered is that the instrument should be stored in an upright position.

What are the different pipetting techniques used?
The frequently used pipetting techniques include the forward pipetting and reverse pipetting. Before we understand these techniques in detail, the general instructions of pipetting listed below would be noteworthy.

  • Press and release the plunger slowly, at all times, particularly when working with high viscosity reagents/solutions. Make sure that the plunger does not snap.
  • Make sure the tip is firmly attached to the tip cone.
  • Before starting your experiment, fill and empty the tip 2-3 times with the reagent or solution that you will be pipetting.
  • Hold the micropipette in an upright position while aspirating. The Grippy must rest on your index finger.
  • Make sure that the tips, the micropipette and the reagent/solution are at the same temperature.

 

Forward Pipetting Technique:

  1. Forward Pipetting TechniqueTo aspirate the liquid in the tip, press the plunger to the first stop. Immerse the pipette tip vertically in the liquid.
  2. Slowly release the plunger while the tip is immersed. The liquid will be aspirated into the pipette tip.
  3. To dispense the liquid, place the tip on the inner wall of the receiving vessel at a steep angle
  4. Slowly press the plunger to the first stop to dispense the liquid.
  5. To empty the tip completely, press the plunger to the second stop.
  6. Wipe the tip on the inner wall while taking the tip out of the vessel.

The Reverse Technique

The reverse technique is suitable for dispensing reagents/solutions that have a high viscosity or a tendency to foam easily. It is also recommended for dispensing very small volumes.

  1. Reverse Pipetting Technique:To aspirate the liquid in the tip, press the plunger to the second stop and immerse the pipette tip vertically in the liquid.
  2. Slowly release the plunger while the tip is immersed. The liquid will be aspirated into the pipette tip.
  3. To dispense the liquid, place the tip on the inner wall of the tube at a steep angle.
  4. Slowly press the plunger to the first stop.
  5. Wipe the tip on the inner wall while taking the tip out of the vessel.

Note: Residual liquid remains in the tip. This does not belong to the dispense volume.

How to choose the right micropipette?
As micropipettes can be used for a number of applications, the dilemma one faces is how to choose the right micropipette while handling a specific task.

For this, you can follow our simple, stepwise guide. The preliminary step includes noting down the details of the experiment you are going to perform with the micropipette. This should ideally include the number of samples, the volume to be transferred, the number of replicates, whether sterile conditions are required and so on. Making a comprehensive list in this manner will make the process of deciding which micropipette to use easier.

Selecting the Type of Micropipette:
Micropipettes can be mainly categorized into two types – Air Displacement Micropipettes and Positive Displacement Micropipettes. You can choose the right one by matching their applications to the experiment you wish to perform.

Next Steps:
After choosing the correct type, let’s look at the sub types available based on criteria like the volume to be handled, quality of tips, manual or electronic pipettes, etc. This is where the more minute details regarding the experiment will come handy.

An adjustable micropipette or a fixed micropipette:
You can either use a micropipette whose volume is already fixed or one whose volume can be adjusted as per requirement. If you regularly use a single volume such as 100 µl for your experiments, go for a fixed volume pipette like MICROLIT RBO Fixed Volume (Single Channel). If your experiment entails you to work with a range of volumes, choose the variable or adjustable pipette, like MICROLIT RBO Variable Volume (Single Channel or Multichannel).

The volume to be handled:
As a standard rule, it is better to choose the smallest pipette capable of handling the required volume because when the set volume is close to the minimum capacity of the micropipette, the accuracy and precision of the readings decrease.

Single channel or multichannel micropipettes:
This decision can be made based on the number of samples or replicates you are working with. Single channel micropipettes like the MICROLIT RBO Single Channel (Fixed Volume and Variable Volume) are used when the number of samples is less whereas a large number of samples or well plates can be easily handled by multichannel micropipettes like the MICROLIT RBO Multichannel (8-channel and 12-channel).

Micropipette Tips:
Once you’ve selected your micropipette, this is the next crucial decision to make. Micropipette Tips can be graduated or not graduated, universal or pipette specific, with filter or without filter, sterile or non-sterile etc. If the experiment requires sterile, aseptic conditions, one should use filtered and sterile tips. Universal tips can be used for a wide range of pipettes. It is also important to match the capacity of the tip with the capacity of the micropipette before use.

Accuracy with specific tolerance:
Micropipettes are designed to operate with accuracies within a few percent (generally <5%) of the intended value for the volume. Make sure your micropipette provides the same accuracy as you need for your sampling.

How to sterilize a micropipette?

You can sterilize your micropipette if it is autoclavable. If you use your micropipette for sensitive and hazardous liquid sampling, the sterilization process is required to clean your micropipette. Check your micropipette manual to know the maximum temperature and time duration for the sterilization process. Refer to the ‘Contamination Cleaning’ section to know how to sterilize your micropipette easily as per dispensed liquid in the sampling.

  • The micropipettes can be sterilized by autoclaving them at 121°C (252°F) and 15 PSI for a duration of 15-20 minutes.
  • No special preparation is needed.
  • You may use steam sterilization bags if needed. After autoclaving, the micropipette must be cooled to room temperature for at least two hours. Before pipetting, make sure that the instrument is dry.
  • It is recommended to check the calibration after every sterilization cycle to achieve the best possible precision and accuracy.

When is micropipette calibration required?
In order to maintain the quality of sampling results, the micropipettes must be in good condition and properly calibrated. Normally, the interval at which a micropipette needs to be calibrated depends on many factors:

  • Frequency of pipette use
  • Type of liquid dispensed by the pipettes
  • Handling & care of pipette
  • Pipette applications that require superior accuracy also demand more frequent calibration
  • Regulations and standards published by organizations such as the FDA and ASTM
  • International provides minimum requirements to ensure the quality of laboratory testing results. Regulations specify that all laboratory instruments used in sampling & productions including micropipettes must be routinely calibrated at suitable intervals.

In the same series, the Clinical and laboratory standards institute (CSLI) has been provided the guidelines for single and multi-channel micropipettes that specifies these micropipettes must be calibrated every 3 to 6 months. And a minimum of two volumes must be tested with 10 replicas at both nominal and lowest settings.

Establishing an appropriate calibration frequency minimizes the chances of incorrect liquid delivery in the laboratory and ensures traceability, accountability and confidence in the results.

FAQs/Troubleshooting:

Which pipette tips can I use with micropipette?
It depends on your micropipette, if your micropipette has a universal tipcone, you can use it with all international tips as per your micropipette volume range.

How can I check whether the micropipette is calibrated?
It is advisable that everyday when you start the pipetting in the lab, check your micropipette for its reading correctness. For this, set the micropipette at the nominal volume and dispense distilled water at least 5 times. Measure the dispensed liquid volume on electronic balance. If the mean of your readings is within the ISO standard limit, the micropipette is calibrated. If not, calibrate your micropipette and check it again.

How often do I need to calibrate the micropipette?
Generally, it depends on how often your micropipette is used for sampling. For normal use of micropipette, it should be checked every 3 or 6 months for calibration. As per ISO 8655 standard, it is recommended to get your micropipette calibrated annually.

How do you set the volume on a micropipette?
A small display with 3 number volume range is given on the body of every micropipette. On a fixed volume micropipette, it is fixed volume range. On variable volume micropipette, you can adjust it within the given volume range as per your required volume sampling. To change the volume range numbers, use the rotational dial given at the top of the micropipette plunger or refer to the manual to know how to change volume of the micropipette.

Why should you avoid touching the micropipette tips?
Touching the micropipette tip can cause the liquid to stick to the outside of the tip. The heat transferred from your hands to the tips can affect the delivery of volumes. It also increases the risk of cross contamination in the laboratory. So, it is recommended to handle the tips only with the micropipette.

What is the largest source of pipetting problems?
Human error is the largest source of the pipetting problems which is caused by touching tips, handling micropipette improperly etc.

What happens if you release the micropipette plunger too fast?
If the micropipette plunger is released too fast, you may face the issue of less draw up and dispensing of liquid.

Information Required to quote on a Micropipette

  1. Single channel or Multichannel?
  2. If Multichannel. How many channels? 8,12,16, 24, 32, 48 or 64 channel?
  3. Do you need a mechanical or electronic pipette?
  4. Do you need require fixed or variable volume?
  5. What volume do you require?

Videos

QUIZ Brochures CAPP Presentation