- Reservation on / off control for fan & UV light
- High quality, quiet fan
- Universal electrical outlet
- An additional pre-filter below work surface helps capture large particles and prolongs HEPA filter life
- Stable laminar air flow, from 0.30m/s to 0.6m/s
- Realtime display air speed, time and the status of HEPA filter
- Cumulative running time of HEPA filter
- Power socket installed inside for easy power access
- Audio and visual alarm for filter failure
- Access password
- Interlocking control between Ultraviolet ray light and interior light
Technical Specifications
Spec/Model | BCV-3S1 | BCV-4S1 | BCV-5S1 | BCV-6S1 |
Num. of operator | one people one side | one people one side | two people one side | three people one side |
Laminar Flow | Vertical | |||
Cleanliness | ISO class 5, ISO-14644-1 | |||
CFU | ≤0.5/Dish*h (90mm petri dish) | |||
Air Velocity |
0.30~0.50 m/s |
|||
Noise | ≤62dB | |||
Vibration | ≤3μm | ≤4μm | ≤3um | ≤4μm |
Illumination | ≥900Lx | |||
Power input | AC220V/50Hz | |||
Peak power | 700VA | 1000VA | 1000VA | 1300VA |
External Dimensions W x D x H(mm) | 920×650×645 | 1220×650×645 | 1520×650×645 | 1820×650×645 |
Internal Work Area W x D x H(mm) | 1040×720×1950 | 1340×720×1950 | 1640×720×1950 | 1940×720×1950 |