Structural Functional Features
- Wide temperature range that can be controlled up to 250°C.
- Improved temperature distribution. (±1.5°C at 100°C)
- 3-point temperature calibration, high temperature accuracy over a wide temperature range.
- Inner stainless steel and aluminum shelves provide excellent corrosion resistance and clean maintenance.
- Protects equipment from over current and over temperature through safety circuits.
- Vacuum line and release line are separated and convenient to use.
- The display is mounted at the top, minimizing floor size requirement.
User Convenience Features
- When a vacuum is formed and maintained, the toughened glass is attached to the gasket, so it is kept confined for long-time repeated use.
- Save and use 3 frequently used temperatures.
- PC connection using RS-232 port and USB port.
- Excellent visibility and viewing angle due to application of 5-inch large LCD.
- Wait On/Off timer. (up to 999 hours 59 minutes)
- Toughened glass viewing window for easy internal viewing.
- Microprocessor PID method for precise temperature control.
- Automatic restart as it was operated after power recovery.
- Monitor via mobile app anytime, anywhere with LC Connected. (mobile monitoring system) (when purchased LC GreenBox)
- Microprocessor PID method for precise temperature control.
Outstanding Safety
- Door structure including buffer system allows for internal pressure release through upper and lower ventilation holes when stable vacuum is formed/maintained and when there is an internal explosion.
- Polycarbonate safety cover provided in case of toughened glass breakage.
- 12 events is recorded in case of door open, error, and etc.
- Top-rated overheating protection system. (registration 10 – 0397583)
- Over temperature limit function.
- Safe product surface, even when operated at upper temperature.
- Malfunction prevented by controller lock function.
- Alarm if set temperature deviation outside of acceptable range.
Vacuum Oven Use Configuration
- It is recommended to use the cold trap bath on the vacuum line to protect the weak vacuum pump against water/solvent or to supply dry air to the sample.
Technical Specifications:
Model | OV4-30S | OV4-65S | |
---|---|---|---|
Control panel(L / cu ft) | 3.5 inch | 3.5 inch | |
Chamber volume (L / cu ft) | 28 / 1 | 65 / 2.3 | |
Temperature | Range (℃ / ℉) |
Amb. +15 to 250
/ Amb. +27 to 482
|
Amb. +15 to 250
/ Amb. +27 to 482
|
Fluctuation (±℃ / ℉ at 100℃) | 0.2 / 0.36 | ||
Variation (±℃ / ℉ at 100℃) | 1.5 / 2.7 | 1.5 / 2.7 | |
Heating time to 100℃ (min.) | 70 | 90 | |
Vacuum gauge range, analog (Mpa)
|
0 to 0.1 | 0 to 0.1 | |
Dimension(W×D×H) | Interior (mm / inch) |
302×305×302
/ 11.9×12×11.9
|
402×405×402
/ 15.8×15.9×15.8
|
Exterior (mm / inch) |
534×522×655
/ 21.1×20.6×25.8
|
636×622×755
/ 25.0×24.5×29.7
|
|
Net weight (Kg / lbs) |
69 / 152.1 | 110 / 242.5 | |
Shelves
|
Quantity of shelves (standard/max.)
|
2 / 3
|
3 / 4
|
Max. Load per shelf (Kg / lbs)
|
20 / 44.1
|
20 / 44.1
|
|
Electrical requirements (230V, 50 / 60Hz, A) | 5.7 | 6.7 | |
Cat. No. | AAH13335K | AAH13345K | |
Electrical requirements (120V, 60Hz, A) | 10.2 | 11.7 | |
Cat. No. | AAH13336U | AAH13346U |
※ The lowest temperature that can be controlled depends on changes in room temperature and sample temperature. Therefore, please contact the distributor of our products before purchasing for technical consultation.
※ Technical data (according to DIN 12880, after 2013)
※ You can add an extra shelf that can be placed at the bottom to experiment with more samples same time, and the value of Variation and Fluctuation may increase slightly.