Vertical floor-standing medical autoclaves with prevacuums and drying (55-175L) – AE-B-MD Series
- Sterilization of non-wrapped solid products, such as clinical instruments, with a drying program optional.
- Sterilization of many wrapped solid products and a lot of porous load, such as wrapped clinical instruments or wrapped hospital textile material, with optional drying program.
- To verify the sterilizer with predefined Bowie-Dick and Vacuum Test programs.
- Sterilization of distilled water and non-medicinal liquid preparations, in open bottles.
Features
- Sterilization chamber made of AISI-316L stainless steel. External housing made of AISI-304 stainless steel.
- Equipment controlled by digital PID microprocessor, with 12 predefined and 38 editable programs, adjustable by time, temperature, number of prevacuums, drying time and and type of cycle (solids or liquids).
- Cycles with fractioned prevacuum (from 1 to 3 prevacuums) and final drying by vacuum pump and heating jacket.
- Immediate production of steam through the built-in steam generator.
- 5” color TFT – LCD touchscreen.
- Includes a single-use water tank for an automatic feed of the sterilization chamber. Optional automatic feed of the water tank.
- Bacteriological filter for air inlet.
Advantages
- Fractioned prevacuum and automatic drying.
- Programs for liquids.
- Test cycles: Vacuum test and Bowie-Dick test.
- Heart probe (optional).
- Inlet in chamber for an external validation probe.
- Programmable auto-start (date and time).
- Storage of more than 100 cycle data.
- USB port for downloading data and updates. Ethernet port for PC connection.
- Optional integrated printer.
Safety
- Safety valve and thermostat.
- Hydraulic lid blocking system while positive pressure exists in the chamber.
- Open lid sensor.
- Water level detector (min./max.) in the water tank.
Special models
- Available models with increased power to 6000W (AE-50-B-MD, AE-75-B-MD and AE-110-B-MD) and 9000W (AE-110-B-MD and AE-150-B-MD). The equipments with increased power reduce the total sterilization cycle time by shortening the heating phase particularly in autoclaves with a high workload.
Technical Specifications
Models | AE-50-B-MD | AE-75-B-MD | AE-110-B-MD | AE-150-B-MD |
---|---|---|---|---|
Chamber volume total/usable | 55/50 L | 79/75 L | 115/110 L | 175/153 L |
Usable chamber dimensions Ø x H | 300 x 710 mm | 400 x 600 mm | 400 x 850 mm | 500 x 760 mm |
Independent clean water tank volume | 9 L | 12 L | 12 L | 20 L |
Exterior dimensions L x D x H | 505 x 580 x 1290 mm | 610 x 700 x 1185 mm | 610 x 700 x 1435 mm | 750 x 820 x 1400 mm |
Loading height | 975 mm | 870 mm | 1120 mm | 1085 mm |
Power | 3200 W | 3200 W | 4500 W | 6000 W |
Gross weight | 110 Kg | 140 Kg | 180 Kg | 265 Kg |
Voltage* | 230V 16A | 230V 16A | 400V 32A | 400V 32A |
Frequency | 50/60 Hz | 50/60 Hz | 50/60 Hz | 50/60 Hz |
Adjustable sterilization temperature | 105-134 ºC | 105-134 ºC | 105-134 ºC | 105-134 ºC |
Adjustable sterilization time | 1-250 min | 1-250 min | 1-250 min | 1-250 min |
Adjustable prevacuum pulses | 1-3 | 1-3 | 1-3 | 1-3 |
Adjustable drying time | 1-360 min | 1-360 min | 1-360 min | 1-360 min |
Max. pressure | 2‚1 Barg | 2‚1 Barg | 2‚1 Barg | 2‚1 Barg |
*Other voltages available under request.